Measurement-computing Analyzer488 Manual de usuario Pagina 61

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 281
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 60
Section 3 Front Panel Operation
3.15
with
ATN
asserted,
and continue
SRQ IFC
RENATN
EOI
LOCATION MESSAGE
Press
[2][ENTER]
# of Matches? 00002
to trigger on the second match
Press
[6][ENTER]
Trig Delay? 0000006
to delay 6 events after
the trigger point
Press
[2][2][ENTER]
Post Count? 00022
to capture 22 events
after the trigger point
Press
[ENTER]
When Completed? Stop
to stop recording when
the trigger sequence is
completed
Press
[FAST] FAST
LED lights
to set handshaking to fast
Press
[ARM] ARM
LED lights
to enable the trigger system
Press
[REC] REC
LED lights,
ARM
LED starts to flash
to start recording
Enter the following controller commands at the PC keyboard:
cmd> output 01;Analyzer488 trigger test
cmd> output 00;Pretrigger data
cmd> output 00;Here !
cmd> output 00;Post trigger data
The
ARM
LED stops flashing when the trigger point is found (
!
). It
remains on until the post trigger sequence is complete. When the
trigger sequence is finished, both the
REC
and
ARM
LED turn off.