Section 4 Serial Controller
4.49
OUTPUT
The
OUTPUT
command sends data to the IEEE 488 bus. The Remote Enable
line (REN) is first asserted. If optional device addresses are specified, those devices
will then be addressed to listen. If addresses are not specified, the Analyzer488 must
already be configured to send data, either as a result of an immediately preceding
OUTPUT
command or as the result of a
SEND
command.
If the character count,
count
, is specified then
count
number of characters
will be sent to the bus devices. Otherwise,
OUTPUT
terminates data transfer upon
detection of a serial CR of LF terminator from the serial input. The serial input
terminator(s) are replaced with the bus output terminator(s) before being sent to the
bus devices.
The IEEE 488 bus output terminators can be modified with the
TERM
command.
Refer to this command description for complete information.
TYPE High Level Controller command
SYNTAX OUTPUT[addr[,addr]][#count];data
addr
is a bus device address. Up to 15 addresses may be specified.
count
is the number of characters to
OUTPUT
.
data
is a string of characters to
OUTPUT
terminated by the serial
terminator(s), unless
count
is specified in which case no
terminator is needed.
RESPONSE None
MODE High Level Controller
BUS STATES REN (if SC), *ATN, data (without
addr
)
REN (if SC), ATNMTA, UNL, LAG, *ATN, data
(with
addr
)
QUERY None
RESPONSE