Section 4 Serial Controller
4.30
ENTER
The
ENTER
command reads data from the IEEE 488 bus.
If a device address (with optional secondary address) is specified, that device
will be addressed to talk. If no address is specified, the Analyzer488 must already be
configured to receive data, either as a result of an immediately preceding
ENTER
and
ADDRESS
command, or as a result of a
SEND
sub-command.
If time-out is enabled, a time-out error will occur if the Analyzer488 does not
receive a data byte within the time-out period.
If character count,
count
, is specified, then exactly
count
characters will be
read from the device. If
count
is not specified,
ENTER
terminates reception on
detection of the first line feed (LF) terminator character. Serial output terminators are
appended after
count
characters are read. The terminator may be overridden by
specifying a different terminator,
term
, in the
ENTER
command.
If a terminator,
term
, option is specified, all CR and LF characters in the input
data are unconditionally discarded. When the specified terminator is detected, it is
discarded and replaced with serial output terminator(s) before control is returned to
the serial host. The optional terminator applies ONLY to the
ENTER
command it is
sent with. The terminator returns to a Line Feed on subsequent
ENTER
commands.
If the
EOI
option is specified, all characters are returned to the host until the
EOI
line is detected. The character sent with
EOI
asserted is also returned followed
by the serial output terminator(s).
This command does not have a query option.
TYPE High Level Controller command
SYNTAX ENTER[addr][#count|term|EOI|;count|;term|;EOI]
addr
is an IEEE 488 bus device address.
count
is the number of characters to
ENTER
.
term
and
EOI
override the normal IEEE bus input LF terminator.