Section 6 Instrument Simulator
6.36
Waveform Terminator Yn
The waveform terminator command,
Yn
, is used to set the terminator character
that will be sent by the Analyzer488 after a waveform is sent. A waveform consists of
256 readings and each reading is one data point in a waveform. When the Analyzer488
is addressed to talk, it will append the waveform delimiter to each waveform.
The
Y0
mode is the most commonly accepted terminator, CR-LF. Command
Y1
reverses the sequence to send LF-CR. Command
Y2,
the default value, sends CR
only and
Y3
sends LF only. The command
Y4
sets the waveform terminator to the
value set by using the User Waveform Terminator command,
Zn
.
SYNTAX:
Y0
CR LF
Y1
LF CR
Y2
CR only (default)
Y3
LF only
Y4
Value set by User Waveform Terminator (
Zn
)
Y?
Return current Waveform Terminator selection
EXAMPLES:
PRINT#1,"OUTPUT10;Y2X"
select Carriage Return terminator
at end of waveform