Vista de pagina 24
Section 2 Getting Started
2.3
Enter the following commands at the keyboard controller prompt (
CMD>
) and
observe the Analyzer488 display:
SRQ IFC
RENATN
EOI
LOCATION MESSAGE
CMD> output 16;F0R0X
BUS ACG13 OD
¼ ¼
CMD> enter 16;eoi
BUS TAG16 50 ¼ ¼
NDCV-000.0156E-3
The display of the Analyzer488 after each command shows the last byte that was
transferred on the bus. Although you only see the last data byte, several bytes were
transferred. To view each transaction at at readable rate, set the handshake speed to
Slow by pressing the [SLOW] key. Set the rate to 3 bytes per second by pressing the
[SHIFT] key followed by the [SLOW] key. Use the curosr keys to select a rate of 3
steps per second then press [ENTER].
Enter the same commands to the keyboard controller. Observe the Anlayzer488
display and you will see the bus activity at 3 bytes per second:
SRQ IFC
RENATN
EOI
LOCATION MESSAGE
CMD> output 16;F0R0X
BUS TAG21 55 ¼ ¼
BUS UNL 3F ¼ ¼
BUS LAG21 30 ¼ ¼
BUS 'F' 46 ¼
BUS '0' 30 ¼
BUS 'R' 52 ¼
BUS '0' 30 ¼
BUS 'X' 58 ¼
BUS CR OD ¼ ¼
BUS ACG13 OD ¼ ¼