CB-7080 & CB-7080D Counter/Timer User's Guide #AAN
2-1
#AAN
Description: Read the counter or frequency value.
Syntax: #AAN[chk](cr)
# A delimiter character
AA 2-character HEX module address, from 00 to FF
N=0 Channel-0 of counter or frequency
N=1 Channel-1 of counter or frequency
[chk] 2-character checksum. If checksum is disabled, no [chk]
(cr) 0x0D
Response:
Valid response [chk](data)(cr)
invalid response No Response
no response Syntax error, communication error or address error.
> A delimiter character indicating a valid response
(data) 8-character data (in HEX format)
[chk] 2-character checksum. If checksum is disabled, no [chk]
(cr)=0x0D
Examples:
Counter 0=0x1E=30 (in decimal):
command: $012(cr)
response: !01500600(cr)
command: #010(cr)
response: >0000001E(cr)
Frequency 1=0x1E Hz = 30 Hz (in decimal):
command: $022(cr)
response: !02510600(cr)
command: #021(cr)
response: >0000001E(cr)