CB-7080 & CB-7080D Counter/Timer User's Guide @AAPN(data)
2-45
@AAPN(data)
Description: Set the preset value of counter. The $AA6 command can reset counter to preset value. Refer to
"Preset value setting" for more information.
Syntax: @AAPN(data)[chk](cr)
@ A delimiter character
AA 2-character HEX module address, from 00 to FF
P
N=0 Counter 0
N=1 Counter 1
(data) 8-character HEX value.
[chk] 2-character checksum. If checksum is disabled, no [chk]
(cr) 0x0D
Response:
Valid response !AA(data)[chk](cr)
invalid response ?AA[chk](cr)
no response Syntax error, communication error or address error
! A delimiter character indicating a valid response
? A delimiter character indicating an invalid response
AA 2-character HEX module address
[chk] 2-character checksum. If checksum is disabled, no [chk]
(cr) 0x0D
Examples:
The preset value of counter 0 is FFFF0000h:
command: @01P0FFFF0000(cr)
response: !01(cr)
The preset value of counter 1 is 0000FFFFh:
command: @02P10000FFFF(cr)
response: !02(cr)