Section 4 General Programming
4.27
Line
210
Cautions that all previous macros and log buffer data
will be destroyed.
Lines
290
-
320
Executes if the user selects 'NO' to the caution.
Line
340
Issues the
FACTORY
command to reset all buffers.
Line
350
Delays to allow the Micro488/EX to become ready for
more commands.
Line
360
Issues the
SAVE
command to clear the
NVRAM
ERROR
condition caused by the
FACTORY
command.
Lines
370
-
470
The main program loop. Outputs to the Micro488/EX
every text line in the input file except those lines which
begin with a comment character (
*
). Comments are
printed to the PC's screen.
Lines
480
-
630
Gets the input file name from the user. If a file extension
is not specified, the extension '
MCR
' is used. If the file
specified does not exist, an error is generated and the
program vectors to Line
610
. Line
630
causes the
program to resume at the Line
520
which request a new
input file.
4.5 Restoring Lost Memory
If, during program development, you notice that the USER heap is shrinking and it
is not due to new
MACRO
definitions or
LOG
ed data, the
FACTORY
command can be
used to restore all memory.
In the unlikely event that the
FACTORY
command does not work, remove the top
cover following the directions and cautions in Section 2. With a small flat blade screw
driver, un-seat the RAM IC (
U103
) from the battery socket (
U100
). Carefully, re-
install the RAM IC.