Appendix B Micro488/EX Error Messages
B.1
Micro488/EX Error Messages
The following error messages are returned if an error condition exists and the
STATUS 1 or STATUS 2 command is executed. The error condition is reset after the
message is sent. Only the most recent error is maintained.
Error No. Error Text and Description
00 OK
01 INVALID ADDRESS
Caused by an invalid address outside the allowable IEEE 488 bus range
of 00 to 30 for primary addresses and 00 to 31 for secondary addresses.
02 INVALID COMMAND
Caused by an unrecognized command or invalid parameter.
03 WRONG MODE
Caused by trying to execute a command not allowed within the present
state of the interface. [eg REMOTE16 as a peripheral]
04 Unassigned - Reserved
05 Unassigned - Reserved
06 NO MACRO
A DOMACRO or READ command was received but the MACRO buffer
specified is empty.
07 MACRO OVERFLOW
No memory is available to allocate as a MACRO buffer.
08 COMMAND OVERFLOW
More than 127 characters were received and interpreted as a command.
09 ADDRESS OVERFLOW
More than 15 primary address/secondary address pairs were received.