Section 4 General Programming
4.14
PERIPHERAL
This event occurs when the Micro488/EX is force from the
Controller Active State (*SCCA) to the Peripheral State
(*SC*CA) by receipt of IFC from the System Controller.
This can be useful in detecting receipt of IFC when in the
*SCCA state. This event is edge sensitive.
CONTROLLER
This event occurs when the Micro488/EX receives control
of the bus and transitions from the Peripheral State
(*SC*CA) to the Controller Active State (*SCCA). This
occurs when the Take Control interface message is
received by the Micro488/EX. This event is edge
sensitive.
TRIGGER
This event occurs when the Micro488/EX, as a Peripheral
(*CA), receives a Group Execute Trigger (GET) command
from the Active Controller. This event is edge sensitive.
When the
MACRO
is executed, the internal status, as read
by the
STATUS
1
command, is cleared.
CLEAR
This event occurs when the Micro488/EX, as a Peripheral
(*CA), receives a Device Clear (DCL) or a Selected
Device Clear (SDC) command from the Active Controller.
This event is edge sensitive. When the
MACRO
is executed,
the internal status, as read by the
STATUS
1
command, is
cleared.
TALK
This event occurs when the Micro488/EX, as a Peripheral
(*CA), detects its My Talk Address (MTA) command
from the Active Controller. It indicates that the controller
has requested information from the Micro488/EX. This
event is edge sensitive.
LISTEN
This event occurs when the Micro488/EX, as a Peripheral
(*CA), detects its My Listen Address (MLA) command
from the Active Controller. It indicates that the controller
has information it wants to send to the Micro488/EX. This
event is edge sensitive.