DAQFlex Software User's Guide Hardware Reference USB-2001-TC
115
USB-2001-TC
Use the components below to set or get device properties.
Component
Supported
Property/Command
Set/Get
Supported Values
AI
CJC/format
Get
CJC/DEGC, CJC/DEGF, CJC/KELVIN
OFFSET
Get
Floating point numeric
RANGE{ch}
Set/Get
BIP73.125E-3V (±0.073125 volts)
BIP146.25E-3V (±0.14625 volts)
SCALE
Set/Get
SENSOR
Set/Get
TC/B, TC/E, TC/J, TC/K, TC/N, TC/R, TC/S, TC/T
SLOPE
Get
Floating point numeric
STATUS
Get
BUSY, ERROR, READY
VALUE
Get
Unsigned integer numeric
DEV
FLASHLED/{n}
Set
0 to 255
FWV
Get
Firmware version
ID
Set/Get
Up to 56 characters
MFGCAL
Get
yyyy-mm-dd HH:MM:SS
MFGCAL{item}
Get
YEAR as yyyy; 20xx
MONTH as mm; 01 to 12
DAY as dd; 01 to 31
HOUR as HH; 01 to 23
MINUTE as MM; 01 to 59
SECOND as SS; 01 to 59
MFGSER
Get
Up to 8 hexadecimal digits
Hardware features
One analog input channel, numbered 0
Supports thermocouple types B, E, J, K, N, R, S, and T
Possible gain ranges:
o ±146.25 mV
o ±73.125 mV
512 bytes of nonvolatile FLASH program memory; used for storing configuration information,
calibration data, and user data.