Measurement-computing WBK Options Manual de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 216
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
Three-Wire Quarter-Bridge Configurations
The three-wire quarter-bridge has only one strain-variable element and requires three bridge completion resistors
(BCRs). The internal half-bridge may be any two matched values, but the remaining resistor must match the
external quarter-bridge value precisely. Three of these values may be installed simultaneously when connected
as shown below; the connector pins determine which resistor is used. With all three values installed, WBK16
can accommodate all three quarter-bridge values without changing the internal resistors.
Three-Wire Quarter-Bridge (+),
Using RA (120-Ohm nominal), B Shunt Resistor
IA
Internal Bridge Completion
RA
RB
RC
RD
RE
RF
RG
RH
1
2
3
-Sense
+Sense
+Excitation
-Excitation
JN01
Switches
accessed
through
software
External
Bridge
DB-9
Input
1
8
9
7
5
4
3
6
2
A 120-ohm resistor and its corresponding
shunt value may be installed as shown.
Three-Wire Quarter-Bridge (+),
Using RC (350-Ohm nominal), D Shunt Resistor
IA
Internal Bridge Completion
RA
RB
RC
RD
RE
RF
RG
RH
1
2
3
-Sense
+Sense
+Excitation
-Excitation
JN01
Switches
accessed
through
software
External
Bridge
DB-9
Input
1
8
9
7
5
4
3
6
2
A 350-ohm resistor and its corresponding
shunt value may be installed as shown.
Three-Wire Quarter-Bridge (+),
Using RE (1-KOhm nominal), F Shunt Resistor
IA
Internal Bridge Completion
RA
RB
RC
RD
RE
RF
RG RH
1
2
3
-Sense
+Sense
+Excitation
-Excitation
JN01
Switches
accessed
through
software
External
Bridge
DB-9
Input
1
8
9
7
5
4
3
6
2
A 1000-ohm (or other value) resistor and
its corresponding shunt value may be
installed as shown.
WBK16, pg. 10 949794 WBK16, Strain-Gage Module
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 215 216

Comentarios a estos manuales

Sin comentarios