Section 3 Command Descriptions
3.15
F3 Number Decimal Equiv
F3 Number Decimal Equiv
000 0 008 8
001 1 009 9
002 2 010 10
003 3 020 20
004 4 100 100
005 5 200 200
006 6 210 210
007 7 255 255
Data received for output to the digital ports must be contained within a prefix (
D
)
and a suffix (
Z
). If the amount of data sent is less than the number of bits programmed as
outputs, the least-significant bits will contain the data sent and the most-significant bits
will be cleared to logic zero. If the data sent is greater than the number of bits
programmed for output or selected by the
Pn
command, the
Digital488
will generate a
conflict error and ignore the entire command string. The
Data Strobe
output is pulse for
approximately 50 microseconds after new data is output on the selected port(s).
When the
Digital488
is addressed to talk (
R0
) it asserts
Inhibit
, reads the data
from all ports, unasserts
Inhibit
and outputs the number of characters determined by the
Gn
and
Pn
commands. Leading zeros are not suppressed and the bus terminators are
appended to the output. After output the
Digital488
must be re-addressed to talk to
perform subsequent reads.
EDR
(
R1
) may also be used to capture data in this format.
Example:
OUTPUT 718;"F3X"
select decimal mode
ENTER 718; A$
read data from the
Digital488
DISP A$
display shows
240;165
OUTPUT 718;D100;200ZX
output
100
&
200
to the
Digital488
ENTER 718; A$
read data from the
Digital488
DISP A$
display shows
100;200