Section 3 Command Descriptions
3.6
Data Dn....Z
The
Data
command outputs up to 40 bits of data to the output ports. The number
of bits which can be sent with the
Data
command is limited by the number of bits
programmed as outputs. For formats
F0
through
F3
, if the amount of data sent is less
than the the number of bits programmed as outputs, the least-significant bits will contain
the data sent and the most-significant bits will be cleared to logic zero. If a single port is
selected with the
Port
command, only eight bits may sent with the
Data
command. The
Data Strobe
output is pulse for approximately 50 microseconds after new data is output
on the selected ports.
For formats
F0
through
F3
, data sent by the controller is contained within a prefix
(
D
) and a suffix (
Z
). In format
F4
, the five bytes immediately following the prefix (
D
) is
interpreted as data and the suffix (
Z
) is not used. For the high speed binary
F5
format,
all bytes received are treated as data and the prefix and suffix are not used. Refer to the
Fn
command for additional details.
Dn...Z
n...
represents the data to be outputted, terminated by
Z
.
(note: in the
F4
mode, the
Z
terminator is not allowed)
Example:
CLEAR 718
reset the
Digital488
OUTPUT 718;"C5P1X"
all ports as output, select port 1
OUTPUT 718;"D55ZX"
send
55
to port 1
ENTER 718; A$
read data from port 1
DISP A$
display shows
55
OUTPUT 718;"P0X"
select all ports
OUTPUT 718;"D1234567890ZX"
send data to all 40 bits
ENTER 718;A$
read data from the Digital488
DISP A$
display shows
1234567890