Note: Specifications are subject to change without notice. 928391 Specifications 7-3
Worst Case Temperature Measuerment Error vs. PDaq3000 Ambient Temperature
with Thermocouple at 0ºC (Excludes Thermocouple Error)
AutoZero Disabled
0
1
2
3
4
5
6
7
8
-30 -20 -10 0 10 20 30 40 50 60 70
Ambient Temperature (ºC)
Temperature Measurement Error (ºC) for Types T, J, K, E,
R, S
J
K
T
E
Voltage Measurement Noise vs. Oversampling
0.01
0.10
1.00
10.00
100.00
1 10 100 1000 10000 100000
Oversampling Rate
Noise (RMS Counts), Typical
100mV
10V
2V
1V
500mV
200mV
Range
Note: does not include noise due to quantizing of returned data