C-18 Program Examples in Quick Basic 879696 TempScan / MultiScan, Appendix C
The following program is ALARM2.BAS in the \EXAMPLES\ subdirectory. It will set up the alarm system
of the MultiScan/1200.
• Lines 1-3: Although not necessary, it is good practice to reset the MultiScan/1200 at the start of your
application by sending the Reset Power-On (
*R) command. Wait a few seconds.
• Lines 4-8: To be certain that the reset was successful and the device is ready, Serial Poll the device
until the proper status is returned.
• Lines 9-10: The first thing necessary to start scanning is to setup the measuring mode. In this case, the
MultiScan/1200 will be setup for line-cycle integration / high-speed multi-channel mode.
• Line 11: The Configure Channels (
C) command will be used once to configure a range of channels
from 1 to 24 as Type T thermocouples, where levels above 18 or below 1 will cause a system alarm.
• Line 12: In addition to causing an internal system alarm state, the alarm conditions can also be
attached to any one of the 32 digital output lines. The Assign Digital Alarm Output (
A) command will
be used to assign input Channels 1 and 2 to digital alarm Output 1 in an ORed fashion.
• Lines 13-14: Program for one Pre-Trigger scan via the Set Counts (
Y) command. Then program the
Trigger (trigger start event) and Stop (trigger stop event) via the Set Trigger Configuration (
T)
command to arm the acquisition.
• Lines 15-36: The Set Digital Outputs (
O) command O? will be used to query the status of the digital
output lines. The alarm state can be detected since digital alarm Output 1 was mapped into input
Channels 1 and 2. This example shows alarm detection handling when two or more channels have to
be muxed into one single digital output bit.
The User Status (U) command U11 will be used to show each individual channel alarm status.