6.2.5 Common Mode Voltage > +/-10V
The PCM-DAS16x/16 will not directly monitor signals with common mode voltages
greater than +/-10V. You will either need to alter the system ground configuration to
reduce the overall common mode voltage, or add isolated signal conditioning between
the source and your board.
6.2.6 Isolated Grounds / Single-Ended Inputs
Single-ended inputs can be used to monitor isolated inputs, though the use of the dif-
ferential mode will increase you system’s noise immunity. The diagram below shows
the recommended connections is this configuration.
25
System with a Large Common Mode Voltage,
Connected to a Differential Input
L
arge common
mode voltage
between signal
source & A/D board
GND
Isolation
Barrier
When the voltage difference
between signal source and
A/D board ground is large
enough so the A/D board’s
common mode range is
exceeded, isolated signal
conditioning must be added.
+
-
Input
Amp
To A/D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
10 K
10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your system’s noise immunity.
Isolated Signal Source
Connected to a Single-Ended Input
I/O
Connector
+
-
Input
Amp
To A/D
LL GND
CH IN
A/D Board
Isolated
signal
source