Measurement-computing CIO-DAS160x/1x Manual de usuario Pagina 25

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 24
BASE ADDRESS +1
A/D 4A/D 5A/D 6A/D 7A/D 8A/D 9A/D 10A/D 11
MSB
01234567
A Read-only register.
On read the most significant A/D byte is read.
6.1.2 A/D DATA & CHANNEL REGISTERS (CIO-DAS1602/16)
BASE ADDRESS
A/D 0
LSB
A/D 1A/D 2A/D 3A/D 4A/D 5A/D 6A/D 7
01234567
A read/write register.
READ
On read, it contains the least significant eight digits of the analog input data.
These eight bits of analog input data must be combined with the eight bits of analog input data in
BASE + 1, to form a complete 16-bit number. The data format is 0 = FS; 65,535 = +FS.
WRITE
Writing any data to the register causes an immediate A/D conversion
.
BASE ADDRESS +1
A/D 8A/D 9A/D 10A/D 11A/D 12A/D 13A/D 14A/D 15
MSB
01234567
A Read-only register.
On read the most significant A/D byte is read
.
6.1.3 CHANNEL MUX SCAN LIMITS REGISTER
BASE ADDRESS +2
CH L0CH L1CH L2CH L3CH H0CH H1CH H2CH H3
01234567
A read and write register.
READ
The current channel scan limits are read as one byte. The high channel number scan limit is in the most
significant four bits. The low channel scan limit is in the least significant four bits.
WRITE
The channel scan limits desired are written as one byte. The high channel number scan limit is in the
most significant four bits. The low channel scan limit is in the least significant four bits.
NOTE
Every write to this register sets the current A/D channel MUX setting to the number in
bits 0-3 and resets the FIFO. See BASE + 8.
21