Measurement-computing CIO-DAS160x/1x Manual de usuario Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
Figure 5-2b. Differential Input Theory
Before moving on to the discussion of grounding and isolation, it is important to explain the concepts of
common mode, and common mode range (CM Range). Common mode voltage is depicted in the diagram
above as Vcm. Though differential inputs measure the voltage between two signals, without (almost)
respect to the either signals voltages relative to ground, there is a limit to how far away from ground
either signal can go. Though the CIO-DAS1600 has differential inputs, it will not measure the difference
between 100V and 101V as 1 Volt (in fact the 100V would destroy the board!). This limitation or
common mode range is depicted graphically in Figure 5-3. The CIO-DAS1600 common mode range is
+/- 10 Volts. Even in differential mode, no input signal can be measured if it is more than 10V from the
boards low level ground (LLGND).
Figure 5-3. Common Mode Range
10
+
-
Input
Amp
To A/D
Differential
Input
LL GND
CH High
CH Low
~
Vs
Vs
Vcm
Common Mode Voltage (Vcm) is ignored
by differential input configuration. However,
note that Vcm + Vs must remain within
the amplifier’s common mode range of ±10V
Vcm = Vg2 - Vg1
gg1 2
+1V
+2V
+3V
+4V
-10V
+5V
-9V
+6V
-8V
+7V
-7V
+8V
-6V
+9V
-5V
+10V
-4V
+11V
-3V
+12V
-2V
+13V
-1V
Gray area represents common mode ran
Both V+ and V- must always remain withi
the common mode range relative to LL G
Vcm
With Vcm= +5VDC,
+Vs must be less than +5V, or the common mode range will be exceeded (>+10V)
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 43 44

Comentarios a estos manuales

Sin comentarios