CB-7013, CB-7033 Users Manual
2.6 $AA1
Description: Perform Zero Calibration
Syntax: $AA1[CHK](cr)
$ Delimiter character.
AA Address of setting module (00 to FF)
1 Command for zero calibration.
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax error or communication error may get no
response.
! Delimiter for valid command.
? Delimiter for invalid command.
AA Address of response module (00 to FF).
Example:
Command: $011 Receive: !01
Preform address 01 zero calibration, return successful.
Command: $021 Receive: ?02
When performing address 02 zero calibration, return was
not enabled before performing calibration command.
Related Command:
Sec2.5 $AA0, Sec. 2.15 ~AAEV
Related Topics:
Sec. 1.8 Calibration