USB-2001-TC User's Guide Specifications
13
Compatible thermocouples
Table 3. Compatible sensor type specifications
Parameter
Specification
Thermocouple type
J: 210 °C to 1200 °C
K: 270 °C to 1372 °C
R: 50 °C to 1768 °C
S: 50 °C to 1768 °C
T: 270 °C to 400 °C
N: 270 °C to 1300 °C
E: 270 °C to 900 °C
B: 0 °C to 1820 °C
Temperature accuracy
The following graphs show the errors for each thermocouple type when connected to the USB-2001-TC. The
graphs display the maximum errors at 25 °C and over the full operating temperature range, and account for
cold-junction compensation errors. The graphs were generated using thermocouples connected to subminiature
connectors of the same type.
The CJC sensor resolution is 0.0625 °C. This is the minimum value of the CJC step width. As such, the reading
may result in a saw tooth curve rather than a square curve as the temperature inside the board changes. This is
the expected behavior.