PCI-DAS6071 User's Guide Specifications
36
Table 4. Relative accuracy specifications - all values are (±)
Range
Relative Accuracy (mV)
Single Point
Averaged (Note 2)
±10V
7.37
1.11
±5V
3.68
0.557
±2.5V
1.84
0.278
±1V
0.737
0.111
±500mV
0.368
0.056
±250mV
0.238
0.032
±100mV
0.111
0.015
±50mV
0.082
0.009
0 to 10V
3.68
0.557
0 to 5V
1.84
0.278
0 to 2V
0.737
0.111
0 to 1V
0.368
0.056
0 to 500mV
0.238
0.032
0 to 200mV
0.111
0.015
0 to 100mV
0.082
0.009
Note 2: Averaged measurements assume dithering and averaging of 100 single-channel readings.
Relative accuracy is defined as the measured deviation from a straight line drawn between measured endpoints
of the transfer function. ADC resolution, noise and front-end non-linearity are included in this measurement.