Appendix C IBERR
Programming Reference Manual C-6 GPIB-488
ESRQ—SRQ in "ON" position
Cause: A wait for an RQS can not occur because the GPIB SRQ line is ON.
Solution: Verify the following:
Ignore the ESRQ until all devices are found.
•Make sure
ibdev is called to open all devices on the bus
capable of asserting SRQ.
Test that each device is unasserting SRQ.
Check the cabling. Make sure that all devices are attached
and the connectors are seated properly.
ETAB—Table problem
Cause: There was a problem with a table used in a FindLstn or FindRQS.
Solution: If the error occurs during a FindLstn, this is a warning
and not an error. Ignore the warning or make the buffer
larger.
If the error occurs during a
FindRQS, then none of the
specified devices are requesting services. Verify that the
device list contains the addresses of all online devices.
EHDL—Invalid handle
Cause: EHDL results when an invalid handle is passed to a function call.
Solution: Do not use a device descriptor in a board function or
vice-versa.
Make sure that the board index passed to the call is valid.