Measurement-computing DAQFlex Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 46
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
DAQFlex Message-based Firmware Specification Analog output calibration messages
16
VALUE
Get the value of the analog input channel being calibrated.
Message
"?AICAL{ch}:VALUE"
Response
"AICAL{ch}:VALUE=value"
ch
The AI channel being calibrated.
value
The AI channel value in counts.
Example
"?AICAL{0}:VALUE"
Analog output calibration messages
The AOCAL component provides messages for calibrating the analog outputs on a USB-1608G Series
and USB-2408 Series device.
Refer to the device-specific information in the Hardware Reference chapter for the component properties
and commands supported by each DAQ device.
AOCAL component calibration properties (USB-1608G Series and USB-2408 Series only)
AIOFFSET, AIRANGE, AIRES, AISLOPE, AIVALUE, LOCK, OFFSET, RES, SLOPE, UNLOCK, VALUE
AIOFFSET
Get the offset for the current AICAL:RANGE setting.
Message
"?AOCAL{ch}:AIOFFSET/HEX"
Response
"AOCAL{ch}:AIOFFSET=0xhex value"
ch
The analog output channel.
hex value
The offset currently set for AICAL:RANGE.
Example
"AOCAL{0}:AIOFFSET=0x84B575AF"
Note
The hex value is used to preserve the precision of the double precision floating point
value.
AIRANGE
Get the analog input range.
Message
"?AOCAL{ch}:AIRANGE"
Response
"AOCAL{ch}:AIRANGE=value"
ch
The analog output channel.
value
BIP10V.
Example
"AOCAL{0}:AIRANGE=BIP10V"
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 45 46

Comentarios a estos manuales

Sin comentarios