DaqBoard/1000 and /2000 Series User’s Manual 959495 Calibration 5-1
Calibration 5
DaqBoard/1000 and /2000 Series boards are factory-calibrated. However, if adjustments are needed they
should be completed in the following order:
1. PGA Input and Output
Offset
2. Sam
p
le/Ho
ld Offset
3. A/D Offset and Gain
4. VDC Voltage Reference
5. DAC0 Full-Scale*
6. DAC1 Full-Scale*
7. DAC2 Full-Scale*
8. DAC3 Full-Scale*
*In regard to the various DaqBoard/1000 and /2000 Series boards, DAC applicability is as follows:
DaqBoard/1000 – DAC0 and DAC1 apply
DaqB
oard/1005 – No DAC
s apply
DaqBoard/2000 and /2000c– DAC0 and DAC1 apply
DaqBoard/2001 and /2001c– DAC0, DAC1, DAC2, and DAC3 apply
DaqBoard/2002 and /2002c– No DACs apply
DaqBoard/2003 and /2003c– DAC0, DAC1, DAC2, and DAC3 apply
DaqBoard/2004 and /2004c– DAC0, DAC1, DAC2, and DAC3 apply
DaqBoard/2005 and /2005c – No DACs apply
A Windows-based program, DaqCal.exe, is used to calibrate Daq systems, including analog expansion
cards. DaqC
al.exe i
s used in conjunction with:
• a 4.5-digit, digital multi-meter
• an adjustable voltage calibrator
• an ambient tem
p
erature meter.
To use the calibration program:
1. Launch DaqCal.
Note:
DaqCal is installed automatically from your data acquisition CD as a part of product
support. This takes place during software installation. DaqCal’s default location is the
IOtech DaqX Software fo
lder, in the Programs group.
2. When DaqCal opens you will be prompted to select your device from a list. After doing so,
simp
ly follow the illustrated on-screen instructions.