Measurement-computing CIO-DAS16 Manual de usuario Pagina 21

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 36
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 20
3.3.7 Isolated Grounds / Differential Inputs
Optimum performance with isolated signal sources is assured with the use of the differential input setting. Figure 3-10 below shows the
recommend connections is this configuration..
Figure 3-10. Isolated Grounds/Differential Inputs
17
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
Signal Source
and A/D Board
Already Isolated.
Already isolated si
g
nal source
and A/D board connected to
a differential input.
GND
These
g
rounds are
electrically isolated.
10 K
10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your systems noise immunity.
Vista de pagina 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 35 36

Comentarios a estos manuales

Sin comentarios