INT4:1 - Interrupt selection
Table 5-2. Interrupt Coding
151111
140111
Not Available1011
120011
111101
100101
Not Available1001
Not Available0001
71110
60110
51010
40010
31100
20100
Not Available1000
DISABLED0000
INTERRUPTINT1INT2INT4INT8
BMDE - Burst Mode Enable. 0 = disable, 1 = enabled. The number of channels in the burst are set in BASE +Ah
register.
TS1:0 - A/D trigger source
Table 5-3. Trigger Source Coding
Internal Paced 11
Rising External Pacer 01
Software TriggerX0
TRIGGER SOURCETS0TS1
To perform conversions in Software Mode, the A/D converter samples at 1 MHz. The maximum delay from the
trigger to the first 1 MHz convert pulse is < 1 uSec, since the trigger enables the 1 MHz pulse train to pass through to
the counter circuit to count 3 pulses which in turn will generate the FIFO write pulse. Therefore, the maximum delay
from the Software Trigger (Base + 0 Write) is 3 uSec.
To perform conversions in Pacer modes, the following sequence must be followed by the program:
The counter (Internal Paced) or DIN0 (External Paced) must be set.
Set the Pacer generator, TS1/0, appropriately.
To initiate the conversions, perform a WRITE to Base + 0 (as in Software Mode) or set an external trigger/gate.
To stop conversions, perform a Base + 9 Write cycle. If using an external trigger, the clearing of the trigger pulse will
end conversions.