1110770 to 1.258
1010550 to 2.54
1100330 to 52
1000110 to 101
R0R1R2R3HEXDECRANGE VGAIN
CONTROL CODES UNI-POLAR AOL
6.4.3 GAIN/RANGES CIO-DAS08-AOM
For those who require the exact ranges and gain codes of the MetraByte DAS08-AO, the CIO-DAS08-AOM is available.
The codes entered in Base + 3 register for the desired range follow:
0111E14+/-0.01500
0011C12+/-0.05100
0101A10+/-0.510
000000+/-51
000188+/-100.5
R0R1R2R3HEXDECRANGE VGAIN
CONTROL CODES BI-POLAR AOM
1111F150 to 0.011000
1011D130 to 0.1100
1101B110 to 110
1001990 to 101
R0R1R2R3HEXDECRANGE VGAIN
CONTROL CODES UNI-POLAR AOM
6.5 COUNTER LOAD & READ REGISTERS
COUNTER 0
BASE ADDRESS + 4 (Read / Write)
D0D1D2D3D4D5D6D7
01234567
COUNTER 1
BASE ADDRESS + 5 (Read / Write)
D0D1D2D3D4D5D6D7
01234567
COUNTER 2
BASE ADDRESS + 6 (Read / Write)
D0D1D2D3D4D5D6D7
01234567
The data in the counter read register, and the action taken on the data in a counter load register, is wholly dependent upon the
control code written to the control register.
The counters are 16-bit counters, each with an 8-bit window, the read / load register. Data is shifted into and out of the 16-bit
counters through these 8-bit windows according to the control byte.
You will need an 8254 data sheet if you want to program the 8254 directly at the register level.